Mini Panel Mounted Controllers |
|
|
Our range of products include Mini Panel Mounted Controllers such as BL 981411 PH Mini Controller, Bl 931700 PH Mini Controller With 4-20 MA Recorder Output, BL 982411 ORP Mini Controller, Bl 932700 ORP Mini Controller With 4-20 Ma Recorder Output, BL 983313 Conductivity Mini Controller, BL 983320 Conductivity Mini Controller and many more items.
|
 | BL 981411 PH Mini Controller | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
The BL 981411 pH controller has been designed for easy, affordable installation in tight spaces, ideal for simple and effective process control. The unit is provided with high impedance pH input and can be used with any pH electrode with standard BNC connector. Measurements are clearly displayed on the LCD, while the status LED indicates operating mode. BL 981411 is also provided with a dosing relay. Selecting acid dosing, will cause the relay to activate when the measurement is higher than the set point. If the basic dosing is selected, the relay is activated when the reading falls below the set point. Set point adjustment (from 0 to 14 pH) and calibration procedures are easily performed with trimmers on the front panel. Users can choose from automatic or manual dosing modes with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. An overtime control system advises users when the relay is active too long, to help prevent overdosing. Order Information: BL 981411-0 (12 VDC) and BL 981411-1 (115/230V) are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.0 to 14.0 pH | | Resolution | | 0.1 pH | | Accuracy @ 25°C | | ±0.1 pH | | Calibration | | manual, with offset trimmer | | Dosing Relay | | max 2A (fuse protected), 250 Vac, 30 Vdc | | Dosing Selection | | acid or alkaline contact open=acid dosage=relay ON if measurement > set pointcontact close=alkaline dosage=relay ON if measurement < set point | | Setpoint | | adjustable from 0.0 to 14.0 pH | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Input Impedance | | 102 Ohm | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | Bl 931700 PH Mini Controller With 4-20 MA Recorder Output | | |
| - It's easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
The BL 931700 mini pH controller has been designed for easy, affordable installation in tight spaces to perform simple yet effective process control. Thanks to its compact size, the BL 931700 mini pH controller can be installed right next to tanks or vats. This versatile controller is ideal for a wide variety of applications, such as textiles, papers, photographic solutions, plating baths, chemicals and water treatment.This BL 931700 is provided with a selectable set point for acid or basic dosage. Measurements are directly taken from a pH electrode in the range from 0 to 14 with a 0.01 pH resolution. Accuracy is ensured by two point calibration, performed manually through trimmers on the front panel. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. The overtime control system advises users when the relay is active for too long, helping to prevent overdosing. In addition, this model features a 4-20 mA analog output for recorder connection. Order Information: BL 931700-0 (12 VDC) and BL 931700-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.00 to 14.00 pH | | Resolution | | 0.01 pH | | Accuracy 25°C | | ±0.02 pH | | Calibration | | manual, 2 point, with offset and slope trimmers | | Dosing Relay | | max 2A (fuse protected), 250 Vac, 30 Vdc | | Dosing Selection | | acid or alkaline contact open=acid dosage=relay ON if measurement > set pointcontact close=alkaline dosage=relay ON if measurement < set point | | Setpoint | | adjustable from 0.00 to 14.00 pH | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Recorder Output | | 4-20 mA, accuracy ±0.20 mA, 500 Ohm max load | | Input Impedance | | 102 Ohm | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 982411 ORP Mini Controller | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
BL 982411 is an ORP mini controller for panel mounting, specially designed for swimming pools and spas. Thanks to its compact size, BL 982411 can be installed in small spaces. BL 982411 is the ideal solution for those who have always checked ORP manually. With it’s automatic dosing, this mini controller will significantly reduce maintenance time. BL 982411 can be used with any ORP electrode with standard BNC connector. The instrument measures ORP in the 0 to 1000 mV range and shows the readings on the display. The status LED continuously indicates if the controller is in measurement, dosing or alarm mode. BL 982411 is also provided with a relay for selecting the dosing direction, oxidizing or reducing. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. The overtime control system advises users when the relay is active for too long, helping to prevent over-dosage. Order Information: BL 982411-0 (12 VDC) and BL 982411-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0 to 1000 mV | | Resolution | | 1 mV | | Accuracy @ 25°C | | ±5 mV | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC | | Dosing Selection | | reducing or oxidizing, selectable on the back panel contact open=reductant dosage=relay ON if measure > set pointcontact close=oxidant dosage=relay ON if measure < set point | | Setpoint | | adjustable from 0 to 1000 mV | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Input Impedance | | 102 Ohm | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | Bl 932700 ORP Mini Controller With 4-20 Ma Recorder Output | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
The BL 932700 is an ORP mini controller that has been designed for easy, affordable installation in tight spaces, ideal for simple yet effective process control. As a result of its compact size, the BL 932700 can be installed right next to tanks or vats. This versatile controller is ideal for many applications, such as ORP monitoring bleaching processes, waste water treatment and swimming pools. BL 932700 permits automatic control of installations previously checked manually. The instrument can be set for reducing or oxidizing dosage. It measures in the ±1000 mV range, with 1 mV resolution. Set point adjustment and calibration are simply performed through trimmers on the front panel. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. The overtime control system advises users when the relay is active too long, helping to prevent over dosage. In addition, this model features a 4-20 mA analog output for recorder connection. Order Information: BL 932700-0 (12 VDC) and BL 932700-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | ±1000 mV | | Resolution | | 1 mV | | Accuracy | | ±5 mV | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC | | Dosing Selection | | reducing or oxidizing, selectable on the back panel contact open=reductant dosage=relay ON if measure > set pointcontact close=oxidant dosage=relay ON if measure < set point | | Setpoint | | adjustable from -1000 to 1000 mV | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Recorder Output | | 4-20 mA, accuracy ±0.20 mA, 500 Ohm max load | | Input Impedance | | 102 Ohm | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 983313 Conductivity Mini Controller | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
This HANNA mini controller has been specially designed for water conditioning and growing applications. Compact in size, it can be mounted in confined spaces or even right next to the vat or barrel containing the chemicals. This meter permits automatic control of installations previously checked manually. EC measurements are shown on the display and the multicolored LED continuously indicates if the mini controller is in measurement, dosing or alarm mode. BL 983313 measures in the 0 to 1999 µS/cm range.
Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. The overtime control system advices users when the relay is active for too long, helping to prevent over dosage. Order Information: BL 983313-0 (12 VDC), BL 983313-1 ( 115/230V), are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0 to 1999 µS/cm | | Resolution | | 1 µS/cm | | Accuracy | | ±2% F.S. | | Probe | | HI 7634-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC contact close when measure > set point | | Setpoint | | adjustable from 0 to 1999 µS/cm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 983320 Conductivity Mini Controller | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
This HANNA mini controller has been specially designed for water conditioning and growing applications. Compact in size, it can be mounted in confined spaces or even right next to the vat or barrel containing the chemicals. This meter permits automatic control of installations previously checked manually. EC measurements are shown on the display and the multicolored LED continuously indicates if the mini controller is in measurement, dosing or alarm mode.BL 983320 measures from 0.0 to 199.9 µS/cm. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. The overtime control system advices users when the relay is active for too long, helping to prevent over dosage. Order Information: BL 983320-0 (12 VDC), BL 983320-1 ( 115/230V), are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.0 to 199.9 µS/cm | | Resolution | | 0.1 µS/cm | | Accuracy @ 25°C | | ±2% F.S. | | Probe | | HI 7634-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC contact close when measure > set point | | Setpoint | | adjustable from 0 to 199.9 µS/cm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 983322 Conductivity Mini Controllers | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
This HANNA mini controller has been specially designed for water conditioning and growing applications. Compact in size, it can be mounted in confined spaces or even right next to the vat or barrel containing the chemicals. This meter permits automatic control of installations previously checked manually. EC measurements are shown on the display and the multicolored LED continuously indicates if the mini controller is in measurement, dosing or alarm mode. BL 983322 from 0.00 to 19.99 µS/cm. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. The overtime control system advices users when the relay is active for too long, helping to prevent over dosage. Order Information: BL 983322-0 (12 VDC) and BL 983322-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.00 to 19.99 µS/cm | | Resolution | | 0.01 µS/cm | | Accuracy | | ±2% F.S. | | Probe | | HI 7634-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC contact close when measure > set point | | Setpoint | | adjustable from 0 to 19.99 µS/cm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | Bl 983317 Conductivity Mini Controller Measuring In MS/CM | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
BL 983317 is a panel mounted mini controller with an output relay that has been designed for easy, affordable installation in tight spaces—ideal for simple yet effective process control. This instrument is provided with automatic compensation for variations in temperature. The probe (not included) is easy to clean and requires very little maintenance. The calibration is performed at one point, through a trimmer. All wiring and connections to external devices are done through the terminals on the rear panel. The multicolor LED continuously indicates if the controller is in measurement, dosing or alarm mode. Users can choose automatic or manual dosing mode by a switch on the front panel. Manual control is particularly useful during maintenance operations because it permits operators to enable or disable the dosing relay according to need. To help prevent overdosing, the overtime control system advises users when the relay is active too long. Order Information: BL 983317-0 (12 VDC), BL 983317-1 ( 115/230V), are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.00 to 10.00 mS/cm | | Resolution | | 0.01 mS/cm | | Accuracy @ 25°C | | ±2% F.S. | | Probe | | HI 7632-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC, contact close when measure < set point | | Setpoint | | adjustable from 0 to 10 mS/cm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 983327 Conductivity Mini Controller Measuring in MS/CM | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
BL 983327 is a panel mounted mini controller with an output relay that has been designed for easy, affordable installation in tight spaces—ideal for simple yet effective process control. This instrument is provided with automatic compensation for variations in temperature. The probe (not included) is easy to clean and requires very little maintenance. The calibration is performed at one point, through a trimmer. All wiring and connections to external devices are done through the terminals on the rear panel. The multicolor LED continuously indicates if the controller is in measurement, dosing or alarm mode. Users can choose automatic or manual dosing mode by a switch on the front panel. Manual control is particularly useful during maintenance operations because it permits operators to enable or disable the dosing relay according to need. To help prevent overdosing, the overtime control system advises users when the relay is active too long. Order Information: BL 983327-0 (12 VDC) and BL 983327-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.00 to 10.00 mS/cm | | Resolution | | 0.01 mS/cm | | Accuracy @ 25°C | | ±2% F.S. | | Probe | | HI 7632-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC, contact close when measure > set point | | Setpoint | | adjustable from 0 to 10 mS/cm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 983315 TDS Mini Controller 0.1 mg/L Resolution | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
This instrument has been designed for TDS control in hydroponics, horticulture and water conditioning. Compact in size, it can be mounted in confined spaces or even right next to the vat or barrel containing the chemicals. These meters permit automatic control of installations previously checked manually. Readings are shown on the display and the multicolored LED continuously indicates if the mini controller is in measurement, dosing or alarm mode. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to your need. The overtime control system advises users when the relay is active too long, helping to prevent over dosage. Order Information: BL 983315-0 (12 VDC), BL 983315-1 ( 115/230V), are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.0 to 199.9 ppm | | Resolution | | 0.1 ppm | | Accuracy 25°C | | ±2% F.S. | | TDS Conversion Factor | | 0.5 | | Probe | | HI 7634-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß=2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC; Contact close when measure: > set point | | Setpoint | | adjustable from 0 to 199.9 ppm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 983319 TDS Mini Controller 1 MG/L Resolution | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
This instrument has been designed for TDS control in hydroponics, horticulture and water conditioning. Compact in size, it can be mounted in confined spaces or even right next to the vat or barrel containing the chemicals. These meters permit automatic control of installations previously checked manually. Readings are shown on the display and the multicolored LED continuously indicates if the mini controller is in measurement, dosing or alarm mode. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to your need. The overtime control system advises users when the relay is active too long, helping to prevent over dosage. Order Information: BL 983319-0 (12 VDC), BL 983319-1 ( 115/230V), are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0 to 1999 ppm | | Resolution | | 1 ppm | | Accuracy 25°C | | ±2% F.S. | | TDS Conversion Factor | | 0.65 | | Probe | | HI 7634-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC Contact close when measure: < set point | | Setpoint | | adjustable from 0 to 1999 ppm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 983321 TDS Mini Controller 0.01 MG/L Resolutio | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
This instrument has been designed for TDS control in hydroponics, horticulture and water conditioning. Compact in size, it can be mounted in confined spaces or even right next to the vat or barrel containing the chemicals. These meters permit automatic control of installations previously checked manually. Readings are shown on the display and the multicolored LED continuously indicates if the mini controller is in measurement, dosing or alarm mode. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to your need. The overtime control system advises users when the relay is active too long, helping to prevent over dosage. Order Information: The BL 983321 is supplied with mounting brackets, transparent cover, and instructions. - Specifications
- Accessories
- Downloads
| Range | | 0 to 19.99 ppm | | Resolution | | 0.01 ppm | | Accuracy 25°C | | ±2% F.S. | | TDS Conversion Factor | | 0.5 | | Probe | | HI 7634-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC Contact close when measure: > set point | | Setpoint | | adjustable from 0 to 19.99 ppm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 983318 TDS Mini Controller 0 to 10,000 ppm | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
BL 983318 is a mini controller that has been designed for easy, affordable installation in tight spaces, ideal for simple yet effective process control. BL 983318 features ATC (Automatic Temperature Compensation) and simple one point calibration performed through the trimmer. The multicolored LED continuously indicates if the controller is in measurement, dosing or alarm mode. Wiring and external device connections are extremely simple to perform through the terminals on the rear of the instrument. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. The overtime control system advises users when the relay is active too long, helping to prevent over-dosage. Order Information: BL 983318-0 (12 VDC) and BL 983318-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.00 to 10.00 ppt (0 to 10,000 ppm) | | Resolution | | 0.1 ppt (10 ppm) | | Accuracy 25°C | | ±2% F.S. | | TDS Conversion Factor | | 0.5 | | Probe | | HI 7632-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC contact close when measure > set point | | Setpoint | | adjustable from 0 to 10 ppt (10,000 ppm) | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | Bl 983324 TDS Mini Controller Measuring Up To 49.9 PPM | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
BL 983324 is a panel mounted TDS controller, designed for easy installation, configuration and maintenance. The meter is provided with a dosing relay, activated when the TDS reading exceeds the set point value. Measurements are compensated for temperature variations and shown on the display automatically. A multicolored LED on the front panel continuously indicates if the mini controller is in measurement, dosing or alarm mode. Wiring and external device connections are extremely simple to perform through the terminals on the rear of the instrument. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. The overtime control system advises operators when the relay is active too long, helping to prevent over dosage. Order Information: BL 983324-0 (12 VDC) and BL 983324-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.0 to 49.9 ppm | | Resolution | | 0.1 ppm | | Accuracy 25°C | | ±2% F.S. | | TDS Conversion Factor | | 0.5 | | Probe | | HI 7634-00 EC/TDS probe with internal temperature sensor and 2 m (6.6’) cable (not included) | | Temperature Compensation | | Automatic, 5 to 50°C (41 to 122°F) with ß = 2%/°C | | Calibration | | manual, with CAL trimmer | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 VDC contact close when measure > set point | | Setpoint | | adjustable from 0 to 49.9 ppm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
 | BL 983314 Resistivity Mini Controller | | |
| - Easy to handle
- Fire retardant casing
- Selectable overdose protection system
- Splash resistant cover
BL 983314 is a simple to operate resistivity controller designed for ultra pure water, reverse osmosis, and water conditioning applications. The BL 983314 resistivity controller is also ideal for continuous monitoring of process solutions. Set point and calibration are manually adjusted with a trimmer and the alarm relay allows for simple control. Readings are automatically temperature compensated, with three different coefficients (β=2.4, 3.5 or 4.5 %/°C). The alarm contact can be used for connection to an alarm, pump, solenoid or dosing system. The relay contact is open when readings are higher than the set point, while for measurements lower than set point, the relay contact is closed. The Hysteresis is typically 0.20 MW/cm from the set point. Measurements are displayed on the LCD and the multicolored LED continuously indicates if the controller is in measurement, dosing or alarm mode. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. The overtime control system advises users when the relay is active for too long, helping to prevent over-dosage. Order Information: BL 983314-0 (12 VDC) and BL 983314-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual. - Specifications
- Accessories
- Downloads
| Range | | 0.00 to 19.90 MΩ/cm | | Resolution | | 0.10 MΩ/cm | | Accuracy @ 25°C | | ±2% F.S. | | Probe | | HI 3314 resistivity probe with 2 m (6.6’) cable (included) | | Temperature Compensation | | Automatic and linear, 5 to 50°C (41 to 122°F) | | Temperature Coefficient | | ß=2.4; 3.5; 4.5%/°C selectable through jumper on the rear panel | | Calibration | | factory calibrated | | Dosing Relay | | maximum 2A (fuse protected), 250 Vac, 30 Vdc contact close when measure < set point | | Setpoint | | adjustable from 0 to 19.90 MΩ/cm | | Overtime | | adjustable, typically from 5 to approx. 30 minutes | | Power Supply | | See Chart | | Dimensions | | 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’) | | Weight | | See Chart | |
|
 Product Showroom View all our products / services as a slide show. play now

|